Table 1 Accuracy (bias), repeatability (CV) and total imprecision (CV) data at the different QC levels for iohexol and creatinine-d3 in VAMS samples (n = 2 × 3) calculated via One-way ANOVA

From: Volumetric absorptive microsampling to measure iohexol and creatinine concentrations for estimation of glomerular filtration rate in cats: aligning animal welfare with practical feasibility

  

Nominal value

(µg/mL)

Accuracy

(bias)

Repeatability

(CV)

Total imprecision

(CV)

Iohexol

LLOQ

2

1.4%

8.0%

8.0%

 

QCL

5

-2.9%

9.4%

9.4%

 

QCM

50

-5.4%

0.9%

7.8%

 

QCH

300

-0.5%

4.2%

7.2%

 

QCL cat

5

1.0%

5.5%

7.5%

 

QCH cat

300

2.5%

5.3%

7.2%

Creatinine-d3

LLOQ

2

-2.2%

8.3%

8.3%

 

QCL

5

-3.3%

9.6%

9.6%

 

QCM

50

-2.6%

2.5%

3.0%

 

QCH

300

0.9%

2.8%

6.6%

 

QCL cat

5

1.1%

5.5%

8.0%

 

QCH cat

300

2.8%

4.3%

6.2%

  1. QCL and QCH prepared in cat blood are indicated in bold. CV coefficient of variation, QC quality control, VAMS volumetric absorptive microsampling