Skip to main content

Table 2 Repeatability and reproducibility analysis of RT-dPCR

From: Development and validation of nanoplate-based RT-dPCR assay for canine respiratory coronavirus detection in various clinical samples

Calculated concentration of CRCoV synthetic oligonucleotide (copies/µL)

Intra-assay variation

(repeatability)

Inter-assay variation

(reproducibility)

Mean

(copies/µL)

SD

CV (%)

Mean

(copies/µL)

SD

CV (%)

3.65 × 104

13883.73

92.75

0.67

12021.87

2399.04

19.46

3.65 × 103

1203.20

24.60

2.00

1129.07

221.51

19.62

3.65 × 102

91.12

5.43

5.95

74.59

9.54

12.79