- Research
- Open access
- Published:
Comparative study of immunoassays, a microelectromechanical systems-based biosensor, and RT-QuIC for the diagnosis of chronic wasting disease in white-tailed deer
BMC Veterinary Research volume 20, Article number: 518 (2024)
Abstract
Background
Chronic wasting disease (CWD) is a fatal transmissible spongiform encephalopathy in cervids. The disease is caused by a pathogenic prion, namely PrPSc. Currently, diagnosis of CWD relies on IHC detection of PrPSc in the obex or retropharyngeal lymph nodes (RPLN) or ELISA screening of obex and RPLN followed by IHC confirmation of positive results. In this study, we assessed the performance characteristics of two immunoassays: CWD Ag-ELISA and TeSeE ELISA, RT-QuIC, and MEMS biosensor via testing 30 CWD + and 30 CWD- white-tailed deer RPLN samples.
Results
Both CWD Ag-ELISA and TeSeE ELISA correctly identified all CWD + and CWD- samples. A greater intra-assay coefficient of variation (CV) in S/P ratios was observed for the TeSeE ELISA (16.52%), compared to CWD Ag-ELISA (9.49%). However, the high CV did not affect the qualitative results of triplicate assays when the corresponding manufacturer’s cutoff was used. The MEMS biosensor not only correctly identified all CWD + and CWD- RPLN samples, but also demonstrated a 100% detection rate for all CWD + samples at dilutions from 10− 0 to 10− 3. Evaluation of RT-QuIC indicated that the rate of false negative reactions decreased from 21.98% at 10− 2 dilution to 0% at 10− 4 and 10− 5 dilutions; and the rate of false positive reactions reduced from 56.42% at 10− 2 dilution to 8.89% and 2.22% at 10− 4 and 10− 5 dilutions, respectively. Based on a stringent threshold of 2 x the first 10 fluorescent readings of each well and a final cutoff of 2/3 positive reactions for each sample, RT-QuIC correctly identified all positive and negative samples at 10− 4 and 10− 5 dilutions. Both MEMS biosensor and RT-QuIC achieved 100% sensitivity and 100% specificity under the experimental conditions described in this study.
Conclusions
The two immunoassays (CWD Ag-ELISA and TeSeE ELISA) performed comparably on white-tailed deer RPLN samples. MEMS biosensor is a reliable portable tool for CWD diagnosis and RT-QuIC can be used for routine testing of CWD if appropriate testing parameters and interpretive criteria are applied.
Background
Chronic wasting disease (CWD) is a fatal transmissible spongiform encephalopathy (TSE) in cervids, such as mule deer (Odocoileus hemionus), white-tailed deer (Odocoileus virginianus), elk (Cervus canadensis), sika deer (Cervus nippon), and moose (Alces alces) [1,2,3]. The causative agent of the disease is a pathogenic prion, designated as PrPSc [4]. Although the emergence of PrPSc remains to be understood, it is known that PrPSc induces conversion of normal cellular prion protein (PrPC) to PrPSc, a misfolded infectious protein that is resistant to proteases and to commonly used disinfection procedures [5,6,7]. Accumulation of PrPSc aggregates in brain tissue results in a neuropathology characterized by spongiform degeneration, neuronal loss, and glial activation which leads to clinical manifestations, including progressive loss of body condition, tremor, nervousness, excessive salivation, and death [8,9,10,11]. Since the first report of CWD in the 1960s, the disease has spread to free-ranging cervids in 32 U.S. states and four Canadian provinces and in captive cervid facilities in 18 U.S. states and three Canadian provinces [12, 13]. The transmission of CWD involves direct contact with cervid tissues and excreta containing the prion and indirect contact with environmental fomites contaminated by PrPSc [2, 14,15,16]. To detect CWD as early as possible in new locations and monitor changes in disease occurrence over time, the Missouri Department of Conservation has implemented stringent CWD control measures, including the CWD hunter harvest initiative, CWD management zone, and CWD surveillance testing (https://mdc.mo.gov/hunting-trapping/species/deer/chronic-wasting-disease/cwd-surveillance).
Several techniques have been used to detect CWD in cervid samples including immunohistochemistry (IHC), antigen capture enzyme linked immunosorbent assay (ELISA) or enzyme immunoassay (EIA), protein misfolding cyclic amplification (PMCA), and real-time quaking-induced conversion (RT-QuIC) [17,18,19,20,21]. Currently, IHC detection of PrPSc amyloid fibrils in the obex of the brain stem or the retropharyngeal lymph node (RPLN) is considered the gold standard and required by the United States Department of Agriculture (USDA) for the Voluntary CWD Certification Program [22]. ELISA is utilized routinely by veterinary laboratories to screen CWD prion in wild and free-range cervids, ELISA positive results are confirmed by IHC. Both IHC and ELISA have high diagnostic sensitivity and specificity for obex and RPLN samples collected from animals in advanced stages of disease [23,24,25,26,27]. CWD ELISA (or EIA) is based on the sandwich ELISA principle which involves immobilizing PrPSc-specific ligand to the wells of an ELISA plate, applying an aliquot of Obex or RPLN tissue homogenate, and adding an enzyme-labeled antibody against PrPSc to form Ab (ligand)-Ag-Ab-enzyme complex, and then an appropriate substrate for color development as an indicator of positive reaction [28]. RT-QuIC takes advantage of the ability of pathogenic prion to convert a recombinant PrP as an assay substrate to a misfolded, aggregated form of rPrP which can be detected by measuring the fluorescence of Thioflavin T incorporated into the newly formed misfolded protein [25, 29, 30]. The formation of PrpSc aggregates during RT-QuIC reaction is monitored in real time by measuring thioflavin-T incorporation into amyloid fibrils [25, 29]. RT-QuIC, a technology that is suitable for high throughput testing and free of generating infectious prions, can be optimized to achieve high accuracy and consistency for CWD diagnosis, especially with samples collected during early presymptomatic stages of CWD [21, 31, 32]. The impedance-based micro-electromechanical systems (MEMS) biosensors have been used to detect various pathogens and toxins in clinical and food samples [33, 34]. Recently, we have demonstrated that a MEMS biosensor is able to detect PrPSc in cervid blood and tissue samples as well as the misfolded proteins in the ELISA positive control [35]. The CWD MEMS biosensor measures the impedance change induced by the binding monoclonal antibody immobilized on the detection electrodes to PrPSc in testing sample [35].
In this study, we evaluated the performance of immunoassays, MEMS biosensor, and RT-QuIC for the detection of PrPSc in hunter-harvested white-tailed deer RPLN samples. Our goal is to develop and validate new technologies for CWD diagnosis.
Materials and methods
Testing samples
As part of Missouri CWD surveillance program, white-tailed deer retropharyngeal lymph nodes (RPLN) were collected from hunter-harvested dead deer and submitted by the Missouri Department of Conservation (MDC) to the University of Missouri Veterinary Medical Diagnostic Laboratory (VMDL) for CWD diagnosis. Sample collection and submission protocols were approved by MDC Science Branch leadership committee. The samples were initially tested using an antigen-capture enzyme immunoassay according to the manufacturer’s instruction (HerdChek CWD Ag-ELISA, IDEXX). ELISA positive results were confirmed by IHC according to the National Veterinary Services Laboratories’ protocol (NVSL-SOP-0878). ELISA negative results were accepted as negative for CWD prion without further analysis by IHC. After diagnostic testing, the samples became VMDL property per VMDL-I-002, and were stored at -80℃ for additional scientific investigations. From the sample repository, 30 CWD + RPLN samples and 30 CWD- RPLN samples were retrieved for the present comparative study of TeSeE ELISA (Bio-Rad Laboratories), CWD Ag-ELISA (IDEXX Laboratories), MEMS biosensor, and RT-QuIC. This research utilized diagnostic samples to evaluate different diagnostic tests and did not involve live animals. The research protocol was approved by the University of Missouri Institutional Biosafety Committee (protocol number: 11721 2.2).
Comparison of two different immunoassays
For TeSeE ELISA, 200 ± 20 mg of RPLN tissue was trimmed using a disposable scalpel and transferred to a 1.5 mL tube containing grinding beads (Bio-Rad Laboratories). The sample was homogenized according to the manufacturer’s instruction (Bio-Rad Laboratories). Approximately 250 µL of the homogenate was incubated with an equal volume of Reagent A containing Proteinase K at 37 ℃ for 10 min followed by adding 250 µL of Reagent B to the tube and centrifuging for 7 min at 15,000 g. The precipitated protein was mixed with 25 µL of Reagent C, incubated for 5 min at 100 ℃, and mixed again via a brief vortex agitation. The purified sample was diluted with 125 µL of Reagent R6. ELISA procedures were performed as follows: transferring 100 µL of the diluted sample to a designated well of the TeSeE ELISA plate (Bio-Rad Laboratories), incubating for 30 min at 37 ℃ followed by three washes with Reagent R2, adding 100 µL of conjugate (Reagent R7) to each well, incubation for 30 min at 4 ℃ followed by five washes with Reagent R2, adding 100 µL of substrate solution, incubating in dark at 20 ℃ for 30 min, adding 100 µL of stop solution, and determining the optical density using a microplate reader (iMark, Bio-Rad Laboratories) equipped with the TeSeE detection software at 450 nm with a reference wavelength of 620 nm.
For CWD Ag-ELISA, 250 ± 50 mg of RPLN was trimmed using a disposable scalpel and transferred to a 1.5 mm ceramic bead tube containing 900 µL of ddH2O. The sample was homogenized using a Bead Mill homogenizer (VWR Life Science) for two cycles of 1 min at a speed of 6.5 m/s with a 10-second dwell halfway through the 1-minute run. ELISA procedures were carried out as follows: mixing 100 µL RPLN homogenate with 25 µL of sample diluent, transferring 100 µL of the mixture to a designated well of the ELISA plate, incubating at 25 ℃ for 1 h, washing the plate six times with 350 µL of 1 x wash solution, adding 100 µL of conditioning buffer to each well followed by incubating at 20 ℃ for 10 min and three washes with 1 x wash buffer 2, adding 100 µL of HRPO-conjugated anti-PrPSc antibody to each well followed by incubating at 20 ℃ for 1 h and five washes with 350 µL wash solution 2, adding 100 µL of TMB substrate to each well followed by incubating at 20 ℃ for 15 min, adding 100 µL of 1 N HCl stop solution, and determining optical density using a microplate reader (ELX800, BioTek Instruments) at 450 nm with a reference wavelength of 650. In addition, serially diluted CWD + RPLN samples and undiluted CWD- RPLN samples were tested in duplicate using the HerdChek CWD Ag-ELISA kit which served as a basis to evaluate MEMS biosensor sensitivity.
Triplicate homogenates of each RPLN were processed and subjected to ELISA analysis along with appropriate positive and negative controls. Qualitative results (positive or negative) were interpreted by following the corresponding manufacturers’ specific instructions. For semi-quantitative analysis, sample/positive control (S/P) ratio for each RPLN sample was calculated using the following formula: (Sample OD – Negative Control OD)/(Positive Control OD – Negative Control OD). Intra-assay coefficient of variation (CV) for each ELISA method was determined as the average CV for the 30 CWD + RPLN samples and individual CV was calculated using the formula CV = δ/µ where δ = standard deviation and µ = mean S/P value of the triplicate tests for each RPLN.
MEMS biosensor design, fabrication, and analysis
The microelectromechanical system (MEMS)-based biosensor was designed with the functions of measuring impedance changes induced by the binding of antigen to antibody coated on the detection electrodes [35]. The fabrication of the sensors was described previously [35]. In brief, the biosensor consisted of a focusing region with a sample inlet and waste outlet, a trapping region, and a detection region with an antibody inlet, an antibody outlet, and a sample outlet. The focusing region was responsible for concentrating prions in the microfluidic channel using two arrays of focusing electrodes wired in a parallel configuration. The unique features of this region included the distinct thick ramp-down electrode pairs along with the tilted thin film fingers that had varying lengths and widths allowing the centration of prions to the centerline of the microchannel while discarding over 90% of fluid in the original sample. The trapping region consisted of gold electroplated and non-parallel electrode pairs surrounding the detection electrodes to maximize prion concentration in the detection region. The trapping electrodes use positive dielectrophoresis (p-DEP) forces to catch prions on the surface of the interdigitated electrode arrays (IDE). The detection region contained a fluidic microchannel with 2 sets of IDE arrays with one set being coated with anti-prion antibody and the other set serving as a negative control. The design of the MEMS-based biosensor is illustrated in Fig. 1.
The MEMS biosensors were used to test serially diluted CWD + RPLN samples (10− 0 to 10− 3) and undiluted CWD- RPLN samples. In brief, anti-prion monoclonal antibody (mAb) (F99/97.6.1, VMRD) was conjugated to a crosslinker, namely sulfosuccinimidyl 6-[3-(2-pyridyldithio) propionamido] hexanoate (sulfo-LC-SPDP) as described previously [35]. The conjugated mAb (100 µL, 2 µg/mL) was loaded into the antibody inlet of the biosensor and incubated at room temperature for 60 min to allow attaching to the detection electrodes. After sucking the fluid containing unbound mAb from the antibody outlet, the impedance was measured for both detection and control electrodes. Then RPLN homogenate (100 µL) was injected into the sample inlet and allowed to enter the focusing region while waste fluid was sucked out through the waste outlet. The concentrated sample was then pushed to the trapping and detection regions using a function generator (Keysight Technologies) with optimized alternating current (AC) voltages at specific frequencies (5 Vp-p at 6 MHz). The biosensor was kept at room temperature for 30 min for prion-mAb binding. After washing the detection channel with dH2O to remove the unbound prions, the impedance was measured again for both detection and control electrodes at frequencies (the rate at which alternative AC current changes direction per second) from 100 Hz to 100,000 Hz using an Impedance Analyzer (Keysight Technologies). The difference between impedance induced by antibody alone and by antibody-antigen interaction was calculated and used to determine the positive or negative status of a sample. The cutoff for a positive response was set as the impedance of all negative samples + 5 SD.
RT-QuIC analysis
RPLN homogenates were serially diluted in PBS from 10− 2 to 10− 5 and analyzed by RT-QuIC. RT-QuIC master mix contained 20 mM Na2HPO4, 320 mM NaI, 1 mM EDTA, 50 µM Thioflavin T (Sigma), and 0.1 mg/ml RT-QuIC rPrP substrate (Syrian Hamster rPrP: 90–231 developed by CWD Evolution and distributed by VMRD, Pullman, WA, USA). RT-QuIC reaction consisting of 95 µL master mix and 3 µL RPLN homogenate was carried out in a 96-well black/clear bottom plate (Greiner Bio-One) that was incubated at 42℃ in a FLUOstar Omega microplate reader (BMG Labtech). RT-QuIC program was set at 1 min rest followed by 1 min shaking at 700 rpm with fluorescence readings every 15 min (450/460 nm excitation, 480 nm emission, bottom read, 20 flashes per well, manual gain 1,800) for 72 h. Data was analyzed using the MARS analytical software (BMG Labtech). Threshold for a positive reaction was established as 2 x mean value of the first 10 fluorescence readings of each well. At a specific dilution, a sample was defined as RT-QuIC + if two out of three reactions were positive. Rate of amyloid formation (RAF) was calculated as the inverse of time (1/h) to threshold (ToT). For the ease of data analysis, 0.001 was assigned as the RAF for a negative reaction. Each run included triplicate sample wells and duplicate positive (ELISA+/IHC + RPLN homogenate) and negative control (ELISA-/IHC- RPLN homogenate) wells.
Sensitivity and specificity
Diagnostic sensitivity was defined as the percentage of CWD + RPLN samples that were correctly identified by RT-QuIC or MEMS biosensor. Diagnostic specificity was defined as the percentage of CWD- RPLN samples that were correctly excluded by RT-QuIC or MEMS biosensor.
Statistical analysis
To evaluate the sensitivity of MEMS biosensor against CWD Ag-ELISA, we compared the qualitative results (positive or negative) for CWD + RPLN samples at the following dilutions: 10− 0, 10− 1, 10− 2, and 10− 3 [36, 37]. We examined CWD positivity of the 30 CWD + RPLN samples for a total of 480 data points (30 samples x 2 repeats x 4 dilutions x 2 methods). Because ELISA results were positive for all CWD + RPLN samples at 10− 0 dilution and negative at 10− 3 dilution, we only estimated ELISA sensitivity at 10− 1 (51 positive) and 10− 2 (16 positive) dilutions using a generalized linear mixed model with a Bernoulli distribution in a Bayesian framework. Response data were sample positivity results, \(\:{y}_{ij}\). Test type, either BioSensor or ELISA, was included as a fixed intercept, \(\:{\alpha\:}_{i}\), and deer identity was included as a random effect centered on zero, \(\:{\beta\:}_{j}\).
Models evaluating sensitivity and S/P ratio results for CWD testing methods all converged based on Rhat values of ≤ 1.1 for estimated parameters and other diagnostics. Models fit to the data was also reasonable (\(\:{BPV}_{sensitivity}=0.374\), \(\:{BPV}_{sp\:ratio}=0.651\))
Results
CWD ELISA and Ag-ELISA
Upon analysis using TeSeE ELISA and CWD Ag-ELISA, all 30 CWD + RPLN samples were positive and 30 CWD- RPLN samples were negative by both methods based on the criteria set by the manufacturers of these kits. Sample over positive control (S/P) values were also examined (Fig. 1). For the 30 CWD + RPLN samples, TeSeE ELISA resulted in greater SP ratios ranging from 0.56 to 3.13 for individual assays with an average SP of all assays for all samples being 2.68 ± 0.64, compared to CWD Ag-ELISA from 0.71 to 1.79 for individual assays with an average SP of 1.47 ± 0.25. For the 30 CWD- RPLN samples, S/P values resulted from both methods were near zero (TeseE ELISA: 0.003 ± 0.004 and CWD Ag-ELISA: 0.003 ± 0.014). Intra-assay variations for each RPLN sample were observed, especially with TeSeE ELISA (Fig. 2). The coefficient of variation (CV) in S/P ratios for the TeSeE ELISA was 16.52% whereas CV for and CWD Ag-ELISA was 9.49%. However, the variations did not affect the qualitative results (CWD positive or negative) as determined using the corresponding manufacturer’s cutoff for positive.
MEMS biosensor performance
The impedance (MΩ) changes induced by representative CWD + and CWD- RPLN samples at various frequencies are shown in Fig. 3. At frequencies from 100 to 10,000 Hz, concentration-dependent impedance changes were detected for the two CWD + RPLN samples. The differences between 10− 0 dilution and 10− 3 dilution were 12- to 17-fold for a strong CWD + sample (Fig. 3A) and 14- to 24-fold for a weak CWD + sample at frequencies (Fig. 3B) from 100 to 10,000 Hz. Meanwhile, a minimum of 11-fold difference between 10− 3 dilution and the control electrodes was detected (Fig. 3A and B) at frequencies from 100 to 10,000 Hz. At 100,000 Hz, impedance change diminished to a baseline level and no difference was detected between sample dilutions (Fig. 3A, B). In contrast to the CWD + RPLN samples, the two CWD- RPLN samples without dilution induced baseline impedance changes at frequencies from 100 to 100,000 Hz (Fig. 3C).
Impedance change induced by representative RPLN samples at various frequencies (102 to 107 Hz). (A): Strong CWD + by IHC. (B): Weak CWD + by IHC. (C): two CWD- samples. CE: control electrodes. CWD + samples at dilutions from 10− 0 to 10− 3 and CWD- samples with no dilution were analyzed by the MEMS biosensor. Data shown are average impedance from two independent runs
Next, the impedance changes triggered by the 30 CWD + RPLN samples at dilutions from 10− 0 to 10− 3 and the CWD- samples without dilution were measured at a frequency of 1000 Hz (Fig. 4). The average impedance for the 30 CWD + samples was 11.15 ± 0.98 MΩ at 10− 0 dilution, 7.70 ± 0.65 MΩ at 10− 1 dilution, 2.74 ± 1.45 MΩ at 10− 2 dilution, and 0.90 ± 0.26 MΩ at 10− 3 dilution. In contrast, the average impedance of the 30 undiluted CWD- samples was 0.32 ± 0.04 MΩ, indicating that the MEMS biosensor analysis was able to distinguish positive and negative samples even when the positive samples were diluted 1,000 times.
The sensitivity of MEMS biosensor in comparison to CWD Ag-ELISA was summarized in Table 1. Statistical analysis indicated that the true positive rate was 100% for MEMS biosensor on CWD + RPLN samples at dilutions from 10− 0 to 10− 3, 100% for CWD Ag-ELISA on samples with no dilution, and 0% for CWD Ag-ELISA for samples at 10− 3 dilution. The estimated positivity rate for CWD Ag-ELISA was 0.849 (95% CRI: 0.134, 1.000) at 10− 1 dilution and 0.266 (95% CRI: 0.000, 0.893) at 10− 2 dilution.
RT-QuIC performance
The percentage of RT-QuIC positive reactions and the average time to threshold (ToT) for all CWD + and CWD- RPLN samples were summarized in Table 2. At 10− 2 sample dilution, 77.78% of RT-QuIC reactions for CWD + RPLN samples were positive while 20% of reactions for CWD- RPLN samples were also positive. Wide variations in ToT were observed among both CWD + and CWD- RPLN samples. When sample dilutions increased to 10− 4 and 10− 5, 100% of the reactions for CWD + RPLN samples were positive and 8.89% and 2.22%, respectively, of the reactions for CWD- RPLN samples were positive. With the CWD- RPLN samples showing false positivity, only one of the triplicate reactions per sample turned positive. At 10− 4 dilution, the average ToT for all CWD + RPLN samples was 24.55 ± 2.52 h while the average ToT for CWD- samples was 50.99 ± 15.05 h with one of the eight positive reactions having a ToT of 25.5 h. At 10− 5 dilution, clear differences in ToT were detected between CWD + RPLN samples (22.52 ± 2.76 h) and CWD- RPLN samples (49.05 ± 6.72 h).
The seeding activities of all samples, expressed as the rate of amyloid formation (RAF, 1/ToT), are presented in Fig. 5. At 10− 2 dilution, wide inter- and intra-sample variations were observed for samples derived from both CWD + and CWD- deer. As sample dilution increased to 10− 5, RAF values became more uniform among samples and individual reactions for the same sample.
Diagnostic sensitivity and specificity of RT-QuIC and MEMS biosensor
With a cutoff of the mean impedance of all negative samples + 5 SD (0.52), the MEMS biosensor demonstrated a sensitivity of 100% on RPLN sample at dilutions from 10− 0 to 10− 3 and a specificity of 100% on undiluted RPLN samples (Fig. 6A). Using a threshold of 2 x mean value of the first 10 readings and a final cutoff of 2/3 positive reactions, the diagnostic sensitivity of RT-QuIC was 93.33% at 10− 2 dilution, 100% at 10− 3 dilution, 100% at 10− 4 dilution, and 100% at 10− 5 dilution. The specificity ranged from 83.33% at 10− 2 dilution, 93.33% at 10− 3 dilution, 96.67% at 10− 4 dilution, and 100% at 10− 5 dilution: indicating that 10− 5 was the ideal dilution for RPLN RT-QuIC under the experimental conditions described in this study (Fig. 6B).
Diagnostic sensitivity and specificity of MEMS biosensor and RT-QuIC. (A): MEMS biosensor analysis of CWD + samples at dilutions from 10− 0 to 10− 3 and CWD- samples with no dilution. (B): RT-QuIC analysis of CWD + and CWD- samples at dilutions from 10− 2 to 10− 5. Sensitivity was defined as the percentage of CWD + RPLN samples (n = 30) that were correctly detected by MEMS biosensor or RT-QuIC. Specificity was defined as the percentage of CWD- RPLN samples (n = 30) that were correctly excluded by MEMS biosensor (only for undiluted samples) or RT-QuIC
Discussion
The diagnosis of cervid CWD relies heavily on IHC and ELISA. These methodologies are suitable for postmortem detection of CWD prion in the brain and RPLN of the affected cervids but may not work well on samples with a low prion load. New technologies, such as RT-QuIC and MEMS biosensor have recently been developed to detect CWD prion. In this study, we compared the performance of two different immunoassays, RT-QuIC and a MEMS biosensor for the detection of CWD prion in RPLN samples.
With the two immunoassays, TeSeE ELISA, designed for diagnosing bovine spongiform encephalopathy (BSE) and scrapie, is approved by the USDA for the detection of CWD in farmed cervids (Bio-Rad Laboratories User Manual). This method includes the following procedures: tissue dissection, homogenization, purification of PrpSc in tissue homogenate, and subsequent detection of prion by ELISA. The CWD Ag-ELISA is an antigen-capture enzyme immunoassay for the detection of PrpSc in white-tailed deer and mule deer as well as elk in the late stage of the disease (IDEXX CWD Ag-ELISA User Manual). CWD Ag-ELISA does not involve protein purification which renders the test to be operationally efficient, a characteristic favored by laboratories that analyze thousands of samples per day. Currently, the CWD Ag-ELISA kit is not approved by USDA for the diagnosis of CWD in farmed cervids, but some states including Missouri allow the use of this kit to screen wild deer populations. To compare the performance of these immunoassays, we tested archived RPLN samples with known CWD status. Our results demonstrated that both TeSeE ELISA and CWD Ag-ELISA kits correctly detected all positive and negative RPLN samples and the subsamples. We also examined the S/P ratios and found that TeSeE kit yielded higher S/P values along with greater intra-assay coefficient of variation, compared to CWD Ag-ELISA kit. Because both immunoassays worked equally well for qualitatively detecting CWD prion in RPLN samples and CWD Ag-ELISA was easy to perform, we relied on CWD Ag-ELISA for subsequent evaluation of MEMS biosensor.
With MEMS biosensor, the optimal conditions, including antibody concentration, cross-linking, and incubation time, have been extensively studied [33,34,35]. Previously, we evaluated the MEMS biosensor using the positive control antigen from the CWD Ag-ELISA kit and a limited number of RPLN samples [35]. In this study, we validated the MEMS biosensor using 30 CWD + and 30 CWD- RPLN samples and assessed its performance based on the initial results generated using ELISA in combination with IHC. Our results indicated that the biosensor correctly identified all CWD + and CWD- samples. Because both MEMS biosensor and immunoassays are based on antibody-antigen binding, we further assessed the sensitivity of the biosensor relative to CWD Ag-ELISA. The 100% detection rate for MEMS biosensor versus 0% detection rate for CWD AG-ELISA in CWD + RPLN samples at 10− 3 dilution confirmed our previous finding that the MEMS biosensor is highly sensitive [35]. It must be noted that ELISA assays were designed to test brain and RPLN samples with no dilution and the comparative study was simply aimed at evaluating the potential suitability of the biosensor for detecting low level PrPSc in RPLN samples. In this study, we did not evaluate MEMS biosensor using serially diluted CWD- RPLN samples because undiluted CWD- RPLN samples induced only baseline impedance changes. Unlike RT-QuIC, MEMS biosensor does not convert PrPC to PrPSc which reasonably preclude the possibility that diluting negative samples would reverse the impedance response triggered by Ab-Ag binding.
With RT-QuIC analysis, we used NaI as the main ionic component in assay buffer as it was shown by our previous work that NaI significantly improved QT-QuIC sensitivity [38]. We used a stringent threshold for reaction positivity (2 x mean value of the first 10 fluorescence readings of each well) to minimize the effect of sample matrix on prion RT-QuIC reaction. We also evaluated effect of sample dilution on RT-QuIC sensitivity and specificity. Our data indicated that the rate of false negative reactions decreased from 21.98% at 10− 2 dilution to 0% at 10− 4 and 10− 5 dilutions and the rate of false positive reaction reduced from 56.42% at 10− 2 dilution to 8.89% at 10− 4 dilution and 2.22% at 10− 5 dilution. With the false positive reactions seen with CWD- RPLN samples at high dilutions, it was necessary to set specific interpretive criteria for RPLN RT-QuIC. Thus, we examined both ToT, also known Ct or ThT, and the ratio of positive to total reactions for each sample. Because one reaction for a CWD- sample had a short ToT (25.5 h) which fell within the ToT range for CWD + RPLN samples at 10− 4 dilution, we did not use ToT as an interpretive criterion. On the other hand, false positivity was detected in only 1/3 reactions at both 10− 4 and 10− 5 dilutions, we selected positivity in 2/3 of reactions as a criterion for a sample to be called as positive. Based on above parameters, RT-QuIC reached 100% sensitivity and specificity on RPLN samples at 10− 4 and 10− 5 dilutions. Although a number of publications have demonstrated that RT-QuIC is suitable for detecting CWD prion in RPLN and other samples [29, 31, 32], assays interpretive criteria have not been established. Our previous data from RT-QuIC analysis of deer platelet samples showed that 10− 2 dilution of the samples produced most reliable results. The findings of this study indicate that testing RPLN at 10− 4 or 10− 5 dilution with triplicate wells can achieve a satisfactory diagnostic outcome. Given that testing each sample at multiple dilutions with greater numbers of replicates is prohibitively expensive for routine diagnostic purposes, we recommend testing RPLN at 10− 4 or 10− 5 with three reactions.
Conclusion
The sensitivity and specificity as well as the performance characteristics of two immunoassays, MEMS biosensor, and RT-QuIC were evaluated by testing 30 CWD + and 30 CWD- RPLN samples from white-tailed deer. TeSeE ELISA yielded greater S/P values along greater intra-assay CV, compared to CWD Ag-ELISA. The MEMS biosensor achieved 100% sensitivity with CWD + RPLN samples at 10− 0 to 10− 3 dilutions and 100% specificity with CWD- RPLN samples without dilution. RT-QuIC also reached a 100% sensitivity and 100% specificity at a sample dilution level of 10− 4 or 10− 5, using the following criteria (1) time to threshold (ToT) for a positive reaction being 2 x the mean value of the first 10 fluorescence readings of a well and (2) cutoff for a sample to be called positive being that 2/3 reactions are positive. Data from this study indicate that MEMS biosensor is a reliable portable tool for CWD diagnosis and RT-QuIC can be used for routine testing of CWD if appropriate testing parameters and interpretive criteria are set.
Data availability
Data are provided within the manuscript and supplementary information files.
Abbreviations
- CWD:
-
Chronic wasting disease
- CV:
-
Coefficient of variation
- ELISA:
-
Enzyme immunoassay
- ELISA:
-
Enzyme-linked immunosorbent assay
- Hz:
-
Hertz
- IDE:
-
Interdigitated electrode array
- IHC:
-
Immunohistochemistry
- mAb:
-
monoclonal antibody
- NaCl:
-
Sodium chloride
- NaI:
-
Sodium iodide
- PBS:
-
Phosphor buffered saline
- PMCA:
-
Protein misfolding cyclic amplification
- PrPC :
-
Prion protein-cellular
- PrPSc :
-
Prion protein-scrapie
- rPrP:
-
Recombinant prion protein
- RAF:
-
Rate of amyloid formation
- RPLN:
-
Retropharyngeal lymph node
- RT-QuIC:
-
Real-time quacking induced conversion
- ToT:
-
Time to threshold
- Vp-p:
-
Peak to peak voltage
- WB:
-
Western blotting
References
Williams ES, Young S. <ArticleTitle Language=“En”>Neuropathology of chronic wasting disease of mule deer (<Emphasis Type=“Italic”>Odocoileus hemionus</Emphasis>) and elk (<Emphasis Type=“Italic”>Cervus elaphus nelsoni</Emphasis>). Vet Pathol. 1993;30:36–45.
Williams ES, Miller MW, Kreeger TJ, Kahn RH, Thorne ET. Chronic wasting disease of deer and elk: A review with recommendations for management. J Wildl Manag. 2002;66:551–63.
Ågren EO, et al. First Detection of chronic wasting disease in Moose (Alces alces) in Sweden. J Wildl Dis. 2021;57. https://doiorg.publicaciones.saludcastillayleon.es/10.7589/JWD-D-20-00141.
Haley NJ, Hoover EA. Chronic wasting disease of cervids: current knowledge and future perspectives. Annu Rev Anim Biosci. 2015;3:305–25.
Sulkowski E. Spontaneous conversion of PrPC to PrPSc. FEBS Lett. 1992;307:129–30. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/0014-5793(92)80750-b.
Yao HL, Han J, Gao JM, Zhang J, Zhang BY, Guo YJ, Nie K, Gao C, Wang XF, Dong XP. Comparative study of the effects of several chemical and physical treatments on the activity of protease resistance and infectivity of scrapie strain 263K. J Vet Med Ser B. 2005;52:437–43. https://doiorg.publicaciones.saludcastillayleon.es/10.1111/j.1439-0450.2005.00897.x.
Madec JY, Vanier A, Dorier A, Bernillon J, Belli P, Baron T. Biochemical properties of protease resistant prion protein PrPSc in natural sheep scrapie. Arch Virol. 1997;142:1603–12. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s007050050183.
Ersdal C, Ulvund MJ, Benestad SL, Tranulis MA. Accumulation of pathogenic prion protein (PrPSc) in nervous and lymphoid tissues of sheep with subclinical scrapie. Vet Pathol. 2003;40:164–74. https://doiorg.publicaciones.saludcastillayleon.es/10.1354/vp.40-2-164.
Fox KA, Jewell JE, Williams ES, Miller MW. Patterns of PrPCWD accumulation during the course of chronic wasting disease infection in orally inoculated mule deer (Odocoileus hemionus). J Gen Virol. 2006;87:3451–61. https://doiorg.publicaciones.saludcastillayleon.es/10.1099/vir.0.81999-0.
Hille MM, Jewell JE, Belden EL. Cellular distribution of the prion protein in palatine tonsils of mule deer (Odocoileus hemionus) and Rocky Mountain elk (Cervus elaphus nelsoni). J Vet Med Sci. 2019;81:1586–96. https://doiorg.publicaciones.saludcastillayleon.es/10.1292/jvms.19-0358. Epub 2019 Sep 24.
Orge L, Lima C, Machado C, Tavares P, Mendonça P, Carvalho P, Silva J, Pinto ML, Bastos E, Pereira JC, Gonçalves-Anjo N, Gama A, Esteves A, Alves A, Matos AC, Seixas F, Silva F, Pires I, Figueira L, Vieira-Pinto M, Sargo R, Pires MDA. Neuropathology of animal prion diseases. Biomolecules. 2021;11:466. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/biom11030466.
Williams ES, Young S. Chronic wasting disease of captive mule deer: a spongiform encephalopathy. J Wildl Dis. 1980;16:89–98. https://doiorg.publicaciones.saludcastillayleon.es/10.7589/0090-3558-16.1.89.
U.S. Geological Survey. Distribution of chronic wasting disease in North America [Internet]. 2023. https://www.usgs.gov/media/images/distribution-chronic-wasting-disease-north-america-0
Mathiason CK, Hays SA, Powers J, Hayes-Klug J, Langenberg J, Dahmes SJ, Osborn DA, Miller KV, Warren RJ, Mason GL, Hoover EA. Infectious prions in pre-clinical deer and transmission of chronic wasting disease solely by environmental exposure. PLoS ONE. 2009;4:e5916.
Henderson DM, Denkers ND, Hoover CE, Garbino N, Mathiason CK, Hoover EA, et al. Longitudinal detection of prion shedding in saliva and urine by chronic wasting disease-infected deer by real-time quaking-induced conversion. J Virol. 2015;89:9338–47.
Kraft CN, Denkers ND, Mathiason CK, Hoover EA. Longitudinal detection of prion shedding in nasal secretions of CWD-infected white-tailed deer. J Gen Virol. 2023;104. https://doiorg.publicaciones.saludcastillayleon.es/10.1099/jgv.0.001825.
Haley NJ, Jürgen AR. Evolution of diagnostic tests for chronic wasting disease, a naturally occurring prion disease of cervids. Pathogens. 2017;6:35.
Peters J, Miller JM, Jenny AL, Peterson TL, Carmichael KP. Immunohistochemical diagnosis of chronic wasting disease in preclinically affected elk from a captive herd. J Vet Diagn Invest. 2000;12:579–82.
Hibler CP, Wilson KL, Spraker TR, Miller MW, Zink RR, DeBuse LL, Andersen E, Scheitzer D, Kennedy JA, Baeten LA, Smeltzer JF, Salman MD, Powers BE. Field validation and assessment of an enzyme-linked immunosorbent assay for detecting chronic wasting disease in Mule Deer (Odocoileus hemionus), White-Tailed Deer (Odocoileus virginianus), and Rocky Mountain Elk (Cervus elaphus nelsoni). J Vet Diagn Invest. 2003;15:311–9.
Kurt TD, Perrott MR, Wilusz CJ, Wilusz J, Supattapone S, Telling GC, Hoover EA. Efficient in vitro amplification of chronic wasting disease PrPRES. J Virol. 2007;81:9605–8.
John TR, Schätzl HM, Gilch S. Early detection of chronic wasting disease prions in urine of pre-symptomatic deer by real-time quaking-induced conversion assay. Prion. 2013;7:253–8.
USDA APHIS | Chronic Wasting Disease (CWD) [Internet]. Chronic Wasting Dis. CWD. 2020. https://www.aphis.usda.gov/aphis/ourfocus/animalhealth/nvap/NVAP-Reference-Guide/Control-and-Eradication/Chronic-Wasting-Disease
Fox KA, Jewell JE, Williams ES, Miller MW. Patterns of PrPCWD accumulation during the course of chronic wasting disease infection in orally inoculated mule deer (Odocoileus hemionus). J Gen Virol. 2006;87:3451–61.
Haley NJ, Mathiason CK, Zabel MD, Telling GC, Hoover EA. Detection of sub-clinical CWD infection in conventional test-negative deer long after oral exposure to urine and feces from CWD + deer. PLoS ONE. 2009;4:e7990.
Holz CL, Darish JR, Straka K, Grosjean N, Bolin S, Kiupel M, Sreevatsan S. Evaluation of real-time quaking-induced conversion, ELISA, and immunohistochemistry for chronic wasting disease diagnosis. Front Vet Sci. 2022;8:824815.
Schwabenlander MD, Rowden GR, Li M, LaSharr KL, Hildebrand EC, Stone S, Seeling DM, Jennelle CS, Cornicelli L, Wolf TM, Carstensen M, Larsen PA. Comparison of chronic wasting disease detection methods and procedures: Implications for free-ranging white-tailed deer (Odocoileus virginianus) surveillance and management. J Wildl Dis. 2022;58. https://doiorg.publicaciones.saludcastillayleon.es/10.7589/JWD-D-21-00033.
Spraker TR, Balachandran A, Zhuang D, O’Rourke KI. Variable patterns of distribution of PrPCWD in the obex and cranial lymphoid tissues of Rocky Mountain elk (Cervus elaphus nelsoni) with subclinical chronic wasting disease. Vet Rec. 2004;155:295–302.
Tabatabaei MS, Ahmed M. Enzyme-linked immunosorbent assay (ELISA). Methods Mol Biol. 2022;2508:115–34. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/978-1-0716-2376-3_10.
Kang HE, Mo Y, Abd Rahim R, Lee HM, Ryou C. Prion diagnosis: application of real-time quaking-induced conversion. Biomed Res Int. 2017;17:5413936. https://doiorg.publicaciones.saludcastillayleon.es/10.1155/2017/5413936.
Khurana R, Coleman C, Ionescu-Zanetti C, Carter SA, Krishna V, Grover RK, Roy R, Singh S. Mechanism of thioflavin T binding to amyloid fibrils. J Struct Biol. 2005;151:229–38. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.jsb.2005.06.006.
Denkers ND, Henderson DM, Mathiason CK, Hoover EA. Enhanced prion detection in biological samples by magnetic particle extraction and real-time quaking-induced conversion. J Gen Virol. 2016;97:2023–9.
Tewari D, Steward D, Fasnacht M, Livengood J. Detection by real-time quaking-induced conversion (RT-QuIC), ELISA, and IHC of chronic wasting disease prion in lymph nodes from Pennsylvania White-Tailed Deer with specific PRNP genotypes. J Vet Diagn Invest. 2012;33:943–8.
Abdullah A, Dastider SG, Jasim I, Shen Z, Yuksek N, Zhang S, Dweik M, Almasri M. Microfluidic based impedance biosensor for pathogens detection in food products. Electrophoresis. 2019;40:508–20. 10.1002.
Muhsin SA, Al-Amidie M, Shen Z, Mlaji Z, Liu J, Abdullah A, El-Dweik M, Zhang S, Almasri M. A microfluidic biosensor for rapid simultaneous detection of waterborne pathogens. Biosens Bioelectron. 2022;203:113993. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.bios.2022.113993.
Muhsin SA, Abdullah A, Kobashigawa E, Al-Amidie M, Russell S, Zhang MZ, Zhang S, Almasri M. A microfluidic biosensor for the diagnosis of chronic wasting disease. Microsyst Nanoeng. 2023;9:104. https://doiorg.publicaciones.saludcastillayleon.es/10.1038/s41378-023-00569-1.
R Core Team. 2023. _R: A Language and Environment for Statistical Computing_. R Foundation for Statistical Computing, Vienna, Austria. https://www.R-project.org/
Kellner K. 2021. _jagsUI: A Wrapper Around ‘rjags’ to Streamline ‘JAGS’ Analyses_. R package version 1.5.2, <https://CRAN.R-project.org/package=jagsUI
Kobashigawa E, Russell S, Zhang MZ, Sinnott EA, Connolly M, Zhang S. RT-QuIC detection of chronic wasting disease prion in platelet samples of white-tailed deer. BMC Vet Res. 2024;20:152. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12917-024-04005-y.
Acknowledgements
The authors would like to thank Missouri Department of Conservation staff for providing RPLN samples and Christy Zhang for editing the manuscript.
Funding
The present study was funded by a grant from Missouri Department of Conservation (00069940).
Author information
Authors and Affiliations
Contributions
EK processed RPLN samples for RT-QuIC and MEMS Biosensor analysis and conducted RT-QuIC analysis. S.A.M and A.A. fabricated the biosensors and performed biosensor testing and analysis of RPLN samples. K.A. performed immunoassays. E.A.S. performed statistical analysis. A.M. designed MEMS biosensor and oversaw biosensor fabrication, testing, and analysis. S.R., M.Z.Z. and S.Z conceived the project. S.Z. coordinated the work by different team members. M.Z.Z. and S.Z. drafted the manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This research did not involve live animals. Retropharyngeal lymph nodes were collected from hunter-harvested dead deer by MDC as part of Misosuri CWD surveillance program. Sample collection and submission protocols were reviewed and approved by MDC Science Committee. Comparative study protocol was approved by the University of Missouri Institutional Biosafety Committee (protocol number: 11721 2.2).
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Kobashigawa, E., Muhsin, S.A., Abdullah, A. et al. Comparative study of immunoassays, a microelectromechanical systems-based biosensor, and RT-QuIC for the diagnosis of chronic wasting disease in white-tailed deer. BMC Vet Res 20, 518 (2024). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12917-024-04351-x
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12917-024-04351-x